Study of Various Parameters that Influence the Content of Exhaled breath Condensate Used in the Diagnosis of Gastroesophageal Reflux Disease

Authors

  • Michal Greguš
  • Pavol Ďurč
  • Júlia Lačná
  • František Foret
  • Petr Kubáň

DOI:

https://doi.org/10.1515/hjic-2018-0007

Keywords:

capillary electrophoresis, exhaled breath condensate, ionic analysis, pH, gastroesophageal reflux disease

Abstract

In this work, various parameters that influence the ionic content and pH of exhaled breath condensate in terms of the noninvasive diagnosis of gastroesophageal reflux disease were studied. Exhaled breath condensate samples were collected using a miniature and inexpensive sampling device. Capillary electrophoresis with contactless conductometric detection was used to monitor the ionic content of exhaled breath condensate. Background electrolyte composed of 20 mM of 2-(N-Morpholino)ethanesulfonic acid, 20 mM of L-Histidine, 2 mM of 18-Crown-6 and 30 µM of cetyltrimethylammonium bromide facilitated the rapid separation of anions and cations, both in less than 2 minutes. The possibility of contamination of the exhaled breath condensate by saliva is discussed in detail. The day-to-day repeatability (n=5) of the ionic content and pH of the exhaled breath condensate was studied and was satisfactory, reflecting mainly the physiological variability.

Author Biographies

  • Michal Greguš
    Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00 Brno, CZECH REPUBLIC; Department of Chemistry, Masaryk University, Kotlářská 267/2, 611 37 Brno, CZECH REPUBLIC
  • Pavol Ďurč
    Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00 Brno, CZECH REPUBLIC; Department of Chemistry, Masaryk University, Kotlářská 267/2, 611 37 Brno, CZECH REPUBLIC
  • Júlia Lačná
    Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00 Brno, CZECH REPUBLIC; Department of Chemistry, Masaryk University, Kotlářská 267/2, 611 37 Brno, CZECH REPUBLIC
  • František Foret
    Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00 Brno, CZECH REPUBLIC; Department of Bioanalytical Instrumentation, Institute of Analytical Chemistry of the Czech Academy of Sciences, v.v.i., Veveří 97, 602 00 Brno, CZECH REPUBLIC
  • Petr Kubáň
    Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00 Brno, CZECH REPUBLIC; Department of Chemistry, Masaryk University, Kotlářská 267/2, 611 37 Brno, CZECH REPUBLIC; Department of Bioanalytical Instrumentation, Institute of Analytical Chemistry of the Czech Academy of Sciences, v.v.i., Veveří 97, 602 00 Brno, CZECH REPUBLIC

Downloads

Published

2018-06-20

How to Cite

Study of Various Parameters that Influence the Content of Exhaled breath Condensate Used in the Diagnosis of Gastroesophageal Reflux Disease . (2018). Hungarian Journal of Industry and Chemistry, 46(1), 29-33. https://doi.org/10.1515/hjic-2018-0007

Most read articles by the same author(s)